Key Features
- Two temperature control modes: Air mode and Baby mode controlled by micro-computer;
- Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
- Set temperature, Air temperature, Baby temperature and Internal battery capacity can be displayedseparately;
- Multiple failure alarm indications;
- ">37℃" temperature set function as optional configuration;
- "Double wall acrylic/polycarbonate glass hood with side door so that the bassinet can be pulled out from it;
- Passive humidification;
- The height of whole unit can be adjusted;
- Oxygen supply system;
- Observation lamp.
Components
Standard configuration
- Main body (Including the Hood, Controller, Bassinet, Internalbattery, and Observation Lamp),
- Oxygen cylinders,
- Oxygen supply system,
- Skin temperature sensor,
- I.V. Pole,
- Mattress,
- AAdjustable stand
Optional configuration
- ">37℃" override temperature mode,
- Disposable skin temperature sensor,
- Ambulance trolley.
Enviromental Temperature
- Operating range: 10℃~30℃(Set temperature should be3℃ higher than the environment temperature)Ambient air velocity: >0.3 m/s,<1.0m/s
Specifications
- AC power requirement: AC220V-230V/50Hzor AC110-120V/50-60Hz, 400VA
- DC Power requirement: DC12V/10A or DC24V/6A
- Control Mode: Air mode and Baby mode controlled by microcomputer
- Air temperature control range: 25℃-37℃(Override mode 37℃-38℃)
- Baby temperature control range: 34℃-37℃(Override mode 37-37.5℃)
- Skin temperature sensor accuracy: ±0.2℃
- The temperature fluctuation range between incubator temper-ature and average incubator temperature (under stable temper-ature condition): ≤1.0℃
- Temperature uniformity: ≤1.5℃(mattress is on the horizontal position)
- Noise inside hood: ≤52 dB(A)
- Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm,power failurealarm and so on.
- Battery continuous working time: 90min (Set temperature is 36℃— Ambient temperature is 15℃)